Description
AOP-1.1DU2.1DGU-7 Trace Oxygen Sensor Datasheet
1. Product Overview
2. Key Features
Ultra-low detection range: 0–10 ppm (custom ranges up to 10,000 ppm available).
Rapid response: T90 response time <7 seconds.
High accuracy: ±2% full-scale (F.S.) precision.
Self-powered operation: No external power supply required.
Superior stability: Monthly drift <1%; linearity ±2% F.S..
Robust build: HDPE housing for durability in harsh environments.
Compact form factor: Space-saving design for easy integration into devices.
3. Technical Specifications
3.1 Electrical Parameters
| Parameter | Value |
|---|---|
| Working Principle | Divided-potential electrochemical |
| Output Current | 1295–700 μA (in ambient air) |
| Measurement Range | 0–10 ppm (customizable: 0–10,000 ppm) |
| Response Time (T90) | <7 s |
| Accuracy | ±2% F.S. |
| Linearity | ±2% F.S. |
| Monthly Drift | <1% |
| Temperature Coefficient | 3.0%/°C |
| Electrical Interface | Gold-plated concentric rings (inner: negative, outer: positive) |
3.2 Mechanical Parameters
| Parameter | Value |
|---|---|
| Housing Material | HDPE (High-Density Polyethylene) |
| Dimensions | Custom (compact, device-integration optimized) |
| Weight | ~5 g |
3.3 Environmental Parameters
| Parameter | Value |
|---|---|
| Operating Temperature | 0–45°C |
| Operating Humidity | 0–99% RH (non-condensing) |
| Operating Pressure | Atmospheric pressure ±20% |
| Expected Lifespan | >2 years (in air) |
| Shelf Life | 12 months (from delivery date) |
4. Application Scope
Industrial Process Control: Nitrogen/hydrogen production, natural gas pipelines, semiconductor manufacturing, petrochemicals.
Medical Industry: Anesthesia gas monitoring, respiratory equipment, medical oxygen analyzers.
Gas Separation: Oxygen monitoring in nitrogen/argon separation systems.
Environmental Monitoring: Emission gas analysis, hazardous gas detection.
Safety Equipment: Oxygen alarms, portable gas detectors, fixed gas monitoring systems.
5. Installation & Operation Guidelines
5.1 Installation
Orientation: Mount vertically with the gas inlet facing downward to prevent liquid ingress.
Wiring: Connect inner gold-plated ring to negative (−), outer ring to positive (+) of the measurement circuit.
Gas Connection: Use 1/8-inch or 3 mm PTFE tubing; maintain flow rate 200–500 mL/min (avoid excessive flow to prevent damage).
5.2 Operation
Warm-Up: Allow 15-minute warm-up after installation for stable readings.
Calibration:
Zero Calibration: Expose to 100% nitrogen (0 ppm O₂) and adjust zero offset.
Span Calibration: Expose to a known concentration (e.g., 10 ppm) and adjust span gain.
Reading: The sensor outputs a current signal proportional to oxygen concentration; convert to ppm via the calibration curve.
6. Safety & Maintenance
6.1 Safety Precautions
Non-explosion-proof: Do not use in flammable/explosive atmospheres.
No liquid contact: Avoid exposure to water, oil, or corrosive gases (e.g., Cl₂, SO₂).
ESD protection: Use anti-static gloves during handling to prevent electrostatic damage.
6.2 Maintenance
Regular Calibration: Perform zero/span calibration every 3–6 months (depending on usage).
Filter Replacement: Replace the inlet filter every 6 months to prevent dust contamination.
Storage: Store in a cool, dry place (10–30°C, 30–70% RH) in sealed packaging; avoid direct sunlight.
7. Troubleshooting
| Issue | Possible Cause | Solution |
|---|---|---|
| No Output | Wiring error or sensor damage | Check wiring; replace sensor if faulty |
| Unstable Readings | Insufficient warm-up or flow fluctuation | Warm up for 15 min; stabilize gas flow |
| Slow Response | Contaminated filter or sensor aging | Replace filter; recalibrate or replace sensor |
| Large Drift | Temperature variation or expired calibration | Compensate for temperature; recalibrate |
8. Manufacturer Information
Brand: ASAIR (Guangzhou Aosong Electronics Co., Ltd.)
Address: No. 10, Kexue Avenue, Science City, Guangzhou, China
Website: www.aosong.com
Support Email: support@aosong.com
Tel: +86-20-3229 8888
Get a Quote