T8030S
May 14, 2026

T8030S

Model T8030S is a Petroleum Product Vapor Pressure Tester (Reid Method),designed in accordance with GB/T 8030 and ASTM D323 standards. It is used to determine the vapor pressure of gasoline, volatile crude oil and other volatile petroleum products. It features high precision, full automation and safe operation.

Description

T8030S.jpg

T8030S(3.jpg

T8030S(2.jpg

Model T8030S is a Petroleum Product Vapor Pressure Tester (Reid Method),designed in accordance with GB/T 8030 and ASTM D323 standards. It is used to determine the vapor pressure of gasoline, volatile crude oil and other volatile petroleum products. It features high precision, full automation and safe operation.

2. Technical Parameters

ParameterSpecification
Power SupplyAC 220V±10%, 50Hz
Rated Power≤1.8kW
Bath Temperature37.8°C (±0.1°C constant)
Pressure Range0~200kPa
Pressure Accuracy±0.5kPa
Test Station1 (standard)
Ambient Temperature10~40°C
Relative Humidity≤85% RH (non-condensing)
Dimensions480×420×380mm
Net Weight~35kg

3. Key Features

  • Standard Compliance: Fully meets GB/T 8030 and ASTM D323 for Reid vapor pressure testing.

  • Precision Control: High-precision PT100 temperature sensor and digital pressure sensor ensure accurate and repeatable results.

  • Automatic Operation: One-key start, automatic temperature stabilization, pressure detection, data hold and printout.

  • Safety Design: Over-temperature, over-pressure and leakage protection; stainless steel pressure chamber.

  • User-Friendly: Large LCD display, real-time temperature/pressure monitoring, and data storage function.

4. Operating Instructions

  1. Preparation: Place the unit on a horizontal surface. Connect power and ground securely. Fill the water bath to the marked level.

  2. Warm-up: Turn on the power. The system will self-check and heat the bath to 37.8°C (stabilize for ≥30min).

  3. Sample Loading: Draw 50mL of cooled sample (0~4°C) into the sample chamber. Assemble and tighten the pressure vessel.

  4. Test Start: Press the START key. The unit will immerse the vessel in the bath and start timing.

  5. Pressure Reading: After 5min, read the pressure value. Rotate the vessel 90° and read again at 7min. Record the final stable reading.

  6. Completion: Press STOP. Release pressure, clean the chamber, and prepare for the next test.

5. Safety Warnings

  • Pressure Hazard: The vessel operates under pressure. Always ensure proper sealing before testing.

  • Temperature Hazard: The bath is at 37.8°C. Avoid prolonged skin contact.

  • Ventilation: Use in a well-ventilated area or fume hood to prevent vapor accumulation.

  • No Unauthorized Modification: Do not disassemble internal components; this may cause safety hazards.

  • Power Off Before Maintenance: Always disconnect power before cleaning or servicing.

6. Maintenance

  • Daily: Clean the sample chamber and seals after use; check water level.

  • Weekly: Verify temperature calibration with a precision thermometer.

  • Monthly: Inspect pressure hoses for leaks; clean the air filter.

  • Quarterly: Calibrate pressure sensor using a standard pressure gauge.

  • Storage: Keep in a dry, dust-free environment; drain the water bath if unused for long periods.