T8019
Description

2. Technical Specifications
| Parameter | Value |
|---|---|
| Power Supply | AC 220V±10%, 50Hz |
| Max Power | ≤2.5kW |
| Evaporation Bath | Aluminum block, φ260×130mm |
| Test Holes | 3 (φ51×70mm each) |
| Working Temp | 160–165°C (±0.1°C accuracy) |
| Air Pressure | 0.035–0.07MPa (adjustable) |
| Air Flow | 1000±100mL/min per jet |
| Ambient Conditions | 0–40°C, ≤85% RH (non-condensing) |
| Dimensions | 550×450×350mm |
3. Key Features
Standard Compliance: Meets GB/T 8019 and ASTM D381 for accurate gum content analysis.
Dual Control: Digital temperature controller (±0.1°C) and independent flowmeters for each jet.
Efficient Testing: 3 simultaneous test positions with uniform heat distribution.
Safe Design: Over-temperature protection, stainless steel bench, and oil-water separator.
Easy Operation: Desktop structure, intuitive controls, and built-in silent air compressor.
4. Operating Instructions
Preparation: Place on a horizontal surface, connect power and air supply. Ground the unit securely.
Warm-up: Turn on the heater; wait for the bath to stabilize at 160–165°C.
Sample Loading: Pipette 50mL of cooled sample into a pre-weighed beaker (φ51×70mm). Insert into a test hole.
Air Adjustment: Set air pressure to 0.035MPa; adjust each jet flow to 1000±100mL/min.
Test Run: Start the timer (30min for gasoline). The instrument maintains temperature and airflow automatically.
Completion: After 30min, stop airflow. Remove beakers, cool in a desiccator, and weigh to calculate gum content.
5. Safety Warnings
High Temperature: Avoid contact with the 160°C bath during operation—risk of burns.
Ventilation: Use in a fume hood or well-ventilated area to prevent vapor accumulation.
No LPG: Do not test liquefied petroleum gas—risk of overpressure.
Disconnect Power: Unplug before cleaning, maintenance, or moving the unit.
Grounding: Ensure proper grounding to prevent electric shock.
6. Maintenance
Daily: Clean beakers and jets after use; check air filters for contamination.
Weekly: Verify temperature calibration with a precision thermometer; adjust if needed.
Monthly: Drain and clean the oil-water separator; inspect hoses for leaks.
Storage: Keep in a dry, dust-free environment; cover to prevent contamination.
Get a Quote