NI-4130
May 15, 2026

NI-4130

NI PXI‑4130 (referred to as NI‑4130) Model: PXI‑4130 | Type: 3U PXI source measurement unit (SMU) + auxiliary power channel | Status: Active, mainstream high-precision DC test module PXI-4130 I. Basic Specification Full Name: NI PXI‑4130 Source Measure Unit (SMU) Form: Single-slot 3U PXI module, integrating two output channels: Channel 1 (SMU main channel): Four-quadrant, isolated SMU (source / sink current, capable of measuring voltage / current) Channel 0 (Utility auxiliary channel): Single-quadrant low-cost power supply Application: Semiconductor / component characterization, LEDs, batteries, sensors, precise DC testing, IV curves, reliability aging Core Advantages: ±20 V / 2 A, 40 W source, 10 W sink, nA level measurement, 4-wire remote sensing, isolated output

Description

NI PXI‑4130 (referred to as NI‑4130)

Model: PXI‑4130 | Type: 3U PXI source measurement unit (SMU) + auxiliary power channel | Status: Active, mainstream high-precision DC test module PXI-4130

I. Basic Specification

Full Name: NI PXI‑4130 Source Measure Unit (SMU)

Form: Single-slot 3U PXI module, integrating two output channels:

Channel 1 (SMU main channel): Four-quadrant, isolated SMU (source / sink current, capable of measuring voltage / current)

Channel 0 (Utility auxiliary channel): Single-quadrant low-cost power supply

Application: Semiconductor / component characterization, LEDs, batteries, sensors, precise DC testing, IV curves, reliability aging

Core Advantages: ±20 V / 2 A, 40 W source, 10 W sink, nA level measurement, 4-wire remote sensing, isolated output

II. Channel Specifications (Core)

✅ Channel 1 (SMU, Four Quadrants, Isolation)

Voltage: ±20 V (5% over-range), Measuring ±20 V

Current range (5 settings, automatic/manual):

200 μA (Resolution 1 nA) 2 mA

20 mA

200 mA

2 A (requiring APS-4100 auxiliary power supply) Power:

Source (I/III Quadrant): 40 W (20 V × 2 A)

Sink (II/IV Quadrant): 10 W (continuous)

Isolation: Channel - Ground ±60 VDC, CAT I

Measurement: 16-bit, 4-wire remote sensing (Remote Sense)

✅ Channel 0 (Utility, single quadrant)

Voltage: 0 ~ +6 V

Current: 1 A (6 W), only I quadrant (source)

Usage: Power supply to DUT, biasing, auxiliary control

III. Accuracy and Resolution (Typical)

Voltage setting resolution: 120 μV

Current setting resolution: 20 μA (2 A range) / 1 nA (200 μA range)

Voltage measurement accuracy: ±0.02% + 200 μV

Current measurement accuracy: ±0.03% + 10 nA (200 μA range)

IV. Power Supply and Heat Dissipation

Host slot power supply: +5 V / 3.3 V, internal power limit: 2 W / 300 mA (2 A range unavailable)

Optional APS-4100 auxiliary power supply: unlock 2 A / 40 W full power, must be equipped

Operating temperature: 0 ~ +55 ℃; >30 ℃ reduced power consumption

V. Software and Drivers

Driver: NI-DCPower, NI-VISA

Support: LabVIEW, LabWindows/CVI, Python, C#, SCPI

System: Windows 10/11, Linux, Real-Time