631-550A
Description

1.1 Description
1.2 Equipment Required
76-814A Bar Phantom
Imaging system (X-ray/fluoroscopy/mammography)
Film or digital image receptor
Ruler (for dimensional verification)
2. Safety Precautions
Warning: This device is for professional use only by qualified medical physics or radiology personnel.
Do not expose to excessive radiation; follow facility radiation safety protocols.
Keep the phantom clean and dry; avoid physical damage to test patterns.
3. Specifications
| Parameter | Value |
|---|---|
| Material | Solid radiopaque/radiolucent polymer |
| Test Patterns | Line pairs (LP/mm), bar targets, resolution grids |
| Dimensions (L×W×H) | 20 cm × 15 cm × 2.5 cm |
| Weight | ~1.2 kg |
| Compatibility | X-ray, fluoroscopy, mammography (2D imaging) |
4. Operating Instructions
4.1 Setup
Place the 76-814A phantom centered on the imaging system’s patient table/cassette holder.
Align the phantom’s reference marks with the system’s central ray.
Set imaging parameters (kVp, mAs, SID) per your facility’s protocol.
4.2 Test Procedures
4.2.1 Intrinsic Resolution
Expose the phantom’s high-resolution line-pair pattern (e.g., 2–20 LP/mm).
Analyze the image to determine the smallest resolvable line pair (record LP/mm value).
4.2.2 Contrast Sensitivity
Image the phantom’s low-contrast bar targets (varying thickness/density).
Identify the thinnest visible bar to assess contrast detection capability.
4.2.3 Uniformity & Artifact Check
Inspect the image for uneven density, noise, or artifacts (e.g., grid lines, scratches).
Ensure the phantom’s background is uniformly exposed.
5. Maintenance & Storage
Cleaning: Wipe the surface with a mild, non-abrasive disinfectant; avoid harsh chemicals.
Storage: Keep in a dry, dust-free case at room temperature (15–30°C).
Inspection: Periodically check for cracks, scratches, or discoloration (replace if damaged).
6. Warranty & Support
Warranty: 1-year limited warranty against defects in materials and workmanship.
Contact: Fluke Biomedical Radiation Management Services | Cleveland, OH | 440.498.2564 | www.flukebiomedical.com/rms