3500/60
Description
Bently Nevada 3500/60 (6‑Channel Temperature Monitor Module) – English Datasheet
Manufacturer: Bently Nevada (Baker Hughes)
Model: 3500/60 (Part No.: 163179‑01, etc.)
Type: 6‑channel temperature monitoring module (RTD/TC input, no recorder outputs)
Application: Turbine/generator bearing/winding temperature monitoring; part of 3500 TSI system (API 670 compliant).
General
Channels: 6 independent, software‑configurable channels
Input Types: RTD (Pt100/Pt1000) or Thermocouple (TC: J, K, E, T)
Function: Conditions inputs, compares to user‑programmable Alert/Danger setpoints
Difference from 3500/61: No analog recorder outputs (3500/61 has 6 recorder outputs)
Power Consumption: 7 W nominal (from 3500 rack backplane)
Input Ranges & Accuracy
RTD (Pt100): −40 °C to +260 °C; ±0.5 °C accuracy
Thermocouples (TC):
Type E: −100 °C to +1000 °C
Type J: −18 °C to +760 °C
Type K: −18 °C to +1370 °C
Type T: −160 °C to +400 °C
Accuracy: ±1.0–1.5 °C (typical)
Input Impedance: >10 MΩ per channel
Scan Rate: 1 second per channel (update)
Alarm & Outputs
Alarm Setpoints: 2 levels (Alert / Danger) per channel, programmable
Relay Outputs: Via 3500 rack alarm relays (shared with system)
Isolation Options:
Non‑isolated (mix RTD/TC)
TC‑isolated (250 Vdc channel‑to‑channel)
Environmental
Operating Temp: −30 °C to +65 °C (−22 °F to +149 °F)
Storage Temp: −40 °C to +85 °C (−40 °F to +185 °F)
Humidity: 5%–95% RH, non‑condensing
Certifications: ATEX, IECEx, UL 508, API 670
Mechanical
Dimensions (H × W × D): 241.3 × 24.4 × 241.8 mm (9.50 × 0.96 × 9.52 in)
Weight: 0.91 kg (2.0 lb)
Mounting: Single‑slot, installs in 3500 series rack
Key Features
✅ 6‑channel mixed RTD/TC monitoring (non‑isolated or TC‑isolated)
✅ High accuracy & stability for critical machinery temperature points
✅ Dual alarm setpoints (Alert/Danger) per channel
✅ API 670 compliant for turbine/generator protection systems



Get a Quote