PR6424/003-030+CON021
Description
I. Model Description PR6424/003-030
PR6424: Φ16mm Eddy Current Sensor
003: 3-meter Shielded Cable
030: Standard Calibration Specification
CON021: Corresponding Special Pre-Amplifier
II. Sensor Technical Parameters
Probe Diameter: 16mm
Installation Thread: M18×1.5
Cable Length: 3m
Linear Measurement Range: 0 - 4.0mm
Sensitivity: 4V/mm
Standard Installation Gap: 3.0mm
Matching Midpoint Voltage: -12V (with CON021)
Frequency Response: 0 - 12kHz
Coil Resistance: 5 - 8Ω
Operating Temperature: -35℃ - +150℃
Protection Grade: IP67
Compatible Material: 42CrMo4 Ferromagnetic Steel
III. Parameters of CON021 Preamplifier
Working Power Supply: -24VDC (range -18 - -30VDC)
Output Voltage Range: -4V - -20V
Power Consumption: ≤2W
Frequency Response: 0 - 10kHz
Built-in Protection: Reverse Connection, Short Circuit, Open Circuit Protection
Installation Environment: Installed in Control Cabinet
IV. Terminal Wiring
CON021 Terminal:
1: -24V
2: Power Ground 0V
3: Signal Output (connected to TSI module)
4: Signal Ground
V. Installation and Commissioning
Initial Gap Setting for Probe: 3.0mm
Power On Adjustment of Gap to Make Preamplifier Output Stable at -12V and Lock
Secure and Tighten the LEMO Plug to Avoid Loosening and Drift
Keep the Cable Away from Power Lines and Frequency Converters, and Ground Shielding Only at One Point on the Cabinet Side
VI. Application
Used for Steam Turbines, Compressors, Fans
Measurements: Axial Vibration, Axial Displacement, Temperature Difference, Eccentricity
Compatible System: MMS6000, AMS6500 TSI System

Get a Quote